Analysis: BD LSR Fortessa

Location: 300 George Street Lab 2331 & Amistad Lab 105; contact: Lesley.Devine@yale.edu

The Fortessas occupy rooms that are specifically designated as BSL2 space for the purpose of sorting and analyzing unfixed human samples obtained from healthy individuals. While gloves are enough personal protective equipment (PPE) when handling non-human samples, it is important to remember that the analysis of unfixed human samples necessitates additional PPE (such as a face shield, to be provided by the user) and specific steps for equipment disinfection that are not required for other cytometers. It is crucial to prioritize personal protection in these rooms. Moreover, human samples from individuals known to be infected with HIV or HepB/C must be fixed before they can be analyzed on any cytometer, including the Fortessas.

UV Laser: 355nm 60 mW
PMTBandpassLong PassFluorochrome
A740/35690LPBUV737
B450/50410LPDAPI, AlexaFluor 350
C379/28 BUV395
Violet Laser: 405nm 100 mW
PMTBandpassLong PassFluorochrome
A710/50690LPBV711
B640/40635LPQdot 655, BV650
C525/50505LPAmCyan, V500, BV510
D450/50 BV421, Pac-Blue, BV450

Please notify FACS Facility Staff if you need to use BV650 on Fortessa Amistad and filter will be adjusted. It is currently set to detect BV605.

Blue Laser: 488nm 50 mW
PMTBandpassLong PassFluorochrome
A685/35670LPPerCP, PerCP-Cy5.5
B515/20505LPFITC, GFP, CFSE
C488/10 SSC
Green Laser: 532nm 150 mW
PMTBandpassLong PassFluorochrome
A780/40735LPPE-Cy7
B710/50685LPPE-Cy5.5
C660/20635LPPE-Cy5
D610/20600LPPE-Texas Red, PI
E582/15 PE
Red Laser: 640 nm 100 mW 
 
PMTBandpassLong PassFluorochrome
A780/60750LPAPC-Cy7, APC-H7, APC-Alexa Fluor 750
B730/45685LPAlexa Fluor 680, Alexa Fluor 700
C670/30 APC, Alexa Fluor 647